Tensile testing machine in the process of daily life operation, there will be many problems, the common are the following:
1, the test machine host power supply is not bright, can not move up and down.
The solution is to check whether the power line of access testing machine is connected normally; Check whether the emergency stop switch is in the state of screwing; Check whether the power supply voltage of access testing machine is normal; Check whether the safety on the socket of the machine is burned off, please remove the spare fuse to install.
3. Information overload occurs in the prompt box after the computer software is online
The solution is to check whether the communication line between computer and testing machine falls off; Check whether the online selection sensor is correctly selected; Check whether the sensor has been bumped during recent tests or keyboard operation; Check whether the software calibration or calibration function is used before the problem occurs; Check whether the calibration value, calibration value, or other information in hardware parameters has been manually changed.
2, the test machine host power supply has electricity but the equipment can not move up and down.
The solution is to check whether it is 15S(time) after the device can not move, because the host boot need self-check, about 15S time; Check whether the upper and lower limit is in the appropriate position, there is a certain running space; Check whether the power supply voltage of access testing machine is normal.
Post time: Jan-26-2022