DRK-DTC Drug Stability Test Chamber(New)

Short Description:

DRK-DTC is based on a scientific method to create a long-term stable temperature and humidity environment for evaluating the expiration date of drugs to create evaluation conditions to meet the accelerated test, long-term test,suitable for the stability inspection of drugs and new drug development.


Product Detail

Product Tags

Overview

The drug stability test chamber is based on a scientific method to create a long-term stable temperature and humidity environment for evaluating the expiration period of drugs to create evaluation conditions to meet the accelerated test, long-term test, high temperature or high temperature of the chemical drug stability test guidelines. The wet test is suitable for the stability inspection of drugs and new drug development in pharmaceutical companies.

Name

Drug stability test chamber(Basic)

Drug stability test chamber(Upgrade)

Model

DRK-DTC-1

DRK-DTC-2

DRK-DTC-3

DRK-DTC-4

DRK-DTC-5

DRK-DTC-6

Temperature range

0~65℃

Temperature fluctuation

±0.2℃

Temperature uniformity

±0.5℃

Humidity range

25~95%RH

25~95%RH(20%~98%by customization)

Humidity deviation

±3%RH

Light intensity

0~6000LX adjustable ≤±500LX,(Ten-level dimming, 600LX per level, precise control of intensity) Test distance 200mm

0~6000LX adjustable ≤±500LX,(Stepless dimming) test distance 200mm

Timing range

With 99 cycles of the program, each cycle is divided into 30 segments, each segment of 1~99 hoursof cyclic steps

Light source board

None
1set

None
1set

None
1set

None
1set

None
1set

None
1set

Temperature and humidity control method

Balanced temperature and humidity control method

Controller

Large touch screen controller

Ultraviolet energy lamp

(Optional) Ultraviolet spectrum range 320~400nm

Standard configuration) UV spectrum range 320~400nm,
Ultraviolet irradiance 0~2 (w/m²)

Cooling system/method

Fully automatic electronic expansion valve refrigeration control system/imported Danfoss compressor

Temperature/humidity sensor

Pt100 platinum resistance/imported German VAISALA humidity sensor

Working temperature

RT+5~30℃

Power supply

AC 220V±10% 50HZ

AC 380V±10% 50HZ

AC 220V±10% 50HZ

Power

1900W

2200W

3200W

4500W

1900W

2200W

Volume

150L

250L

500L

1000L

150L

250L

WxDxH
Liner size(mm)

480*400*780

580*500*850

800*700*900

1050*590*1610

480*400*780

580*500*850

WxDxH
Dimensions(mm)

670*775*1450

770*875*1550

1000*1100*1860

1410*890*1950

670*775*1450

770*875*1550

Loading tray(Standard)

2pcs

3pcs

4pcs

2pcs

3pcs

Embedded printer

Standard configuration

Safety equipment

Compressor overheating protection, fan overheating protection, over temperature protection, compressor over pressure protection, overload protection, water shortage protection.

Standard

 According to the 2015 edition of Pharmacopoeia drug stability test guidelines and GB/10586-2006 related manufacturing clauses     

Temperature and humidity operation chart:

2

Features

Numbering

Content and description

Standard

URS1

Equipped with large touch control screen, touch screen≥7 inch.

Real-time monitoring of operating conditions, can display current temperature (humidity), temperature (humidity) set value, date, time, temperature (humidity) curve and other working parameters.

The operating parameters can be adjusted arbitrarily.

Yes

URS2

With data storage function, it can store 100,000 data.

Yes

URS3

With the user authority classification function, it can be divided into two user levels: technologist and operator.

Operator authority: view interface information, alarm and data curve functions.

Technician authority: including operator authority, setting process parameters, area interface operation function, start and stop preset program, report query, operation record event query.

Each account must log in with a password before performing operations within the scope of authority.

Yes

URS4

Equipped with an intelligent defrosting system, which does not affect equipment performance during defrosting.

Yes

URS5

The equipment is equipped with a micro printer (printing interval 0~9999 minutes).

Yes

URS6

The equipment has the main functions of heating, humidification, gating, lighting, sterilization, defrosting, and alarm.

Yes

URS7

Equipment operation mode is divided into: fixed value mode and program mode (program mode can be set for 30 segments and 99 cycles).

Yes

URS8

Equipment timing mode: running timing, constant temperature timing, constant humidity timing, constant temperature and humidity timing can be selected.

Yes

URS9

With alarm functions: temperature alarm, humidity alarm, water shortage alarm, door open alarm, etc.

Yes

URS10

Schedule switch machine function.

Yes

URS11

Power-off start function: No start: After power-off and restart, the system is in a stopped state.Hard start: After power off and restart, the system starts to run from the first segment of the first cycle, and the timing time is cleared.Soft start: After power off and restart, the system starts to run from the period of time when the power is off.The three startup modes can be switched freely, and the factory defaults to not start.

Yes

URS12

Standard USB interface, data can be exported instantly

Yes