Product Description
The walk-in drug stability laboratory is designed and manufactured according to user requirements and with reference to GB/T10586-2006, GB/T10592-2008, GB4208-2008, GB4793.1-2007 and other relevant provisions. It is mainly used to manufacture and maintain a space with constant temperature and humidity. The cooling and heating of the device are fully automatically controlled.
Main features:
Brand-new perfect shape design, thickness of 100MM polyurethane insulation warehouse board, external steel plate baking paint, internal SUS#304 stainless steel, internal safety door, internal alarm switch and independent over-temperature alarm system to ensure the safety of operators;
The central control system adopts a Japanese-imported Youyi control touch screen temperature and humidity controller with high temperature control accuracy. It is equipped with USB interface, LAN network cable interface, computer control software to realize remote monitoring, temperature and humidity curve view, data storage, data printer, faulty mobile phone text messages Alarm and other functions;
Control signal acquisition adopts Austrian E+E original imported temperature and humidity transmitter;
The balance temperature control system (BTHC) controls the SSR in a P.I.D. continuous and automatic adjustable manner, so that the heating capacity of the system is equal to the heat loss, so it can be used stably for a long time;
Provide 3Q certification scheme: can provide customers with a series of services such as IQ (installation confirmation), OQ (operation confirmation), PQ (performance confirmation), etc.
The shelves are stainless steel and chrome-plated, and the grille-type laminates can be adjusted.
The main technical parameters:
Specification Model: DRK637
Temperature Range: 15℃∼50℃
Humidity range: 50%RH ∼ 85%RH
Temperature and humidity resolution: Temperature: 0.1℃; Humidity: 0.1%
Carton size: Width 2700 × depth 5600 × height 2200mm
Internal dimensions: Width 2700×Depth 5000×Height 2200mm
Refrigeration system: Adopting Emerson Copeland Scroll Hermetic Compressor, two sets of refrigeration systems, one standby and one use
Cooling Method: Air-cooled
Power: 20KW
Air Conditioning System
Air conditioning method: forced ventilation internal circulation, balanced temperature regulation (BTHC), this method in the case of continuous operation of the refrigeration system, the central control system according to the collected temperature signal inside the box for amplification, analog, digital conversion, non-linear After calibration, it is compared with the set value (target value) of the temperature, and the deviation signal obtained is subjected to PID calculation, and the adjustment signal is output, and the output power of the heater is automatically controlled, and finally the temperature in the box reaches a dynamic balance.
Indoor air circulation device: stainless steel centrifugal fan and air guide plate ensure uniform air supply, uniform indoor temperature and adjustable indoor wind speed.
Air heating method: finned radiator tube high-quality nickel-chromium alloy electric heater heating.
Air cooling method: multi-stage finned air heat exchanger.
Refrigeration method: Two sets of Emerson Copeland Scroll fully enclosed compressors, one for use and one for preparation, environmentally friendly refrigerant R404A.
Humidification method: electrode type humidification.
Dehumidification method: high-efficiency surface condensing dehumidifier.
Central Controller:
Youyi Control 7.0 inches LCD touch controller imported from Japan, Chinese operation interface, man-machine interface touch screen input, fixed value or program action status display, only need to set the temperature and time corresponding parameters during program setting, no additional input compression is required The controller has a data storage function, and can be directly exported by U disk or displayed and printed test data and curves through special software on the PC. Control temperature and humidity setting test value display: It can display the execution program number, segment number, remaining time and cycle times, running time display program editing and graphic curve display, with real-time display program curve execution function.
The main performance indicators of the controller:
Setting method: Programmable and fixed value setting
Memory Capacity: 1000 program memory, each group of 100 steps 999 cycles, 10 groups of program link function
Predetermined Area: Temperature condition: PT100: -100~200℃, humidity condition: 0-100%RH
Display Range: Temperature conditions: PT100_1:–100~200℃, humidity conditions: 0-100%RH
Cumulative Running Time: 99999 hours 59 minutes
Set Resolution: Temperature: ±0.1℃, humidity: ±0.1%RH
Time resolution: 1 minute
Display Resolution: Temperature: ±0.1℃; Humidity: ±0.1%RH
Input Signal: PT(100Ω); DC input power: temperature: 4-20mA humidity: 4-20mA
Control mode: 9 groups of PID control and Fuzzy control
Set Slope Setting: Temperature 0~100℃ per minute
Data Storage Capacity: Can store 600 days of data and curves (1 time/minute)
Operation Setting: Power-off memory can be set, and the last result will continue to run after power-on;
The equipment can be started and shut down by appointment; the date and time can be adjusted;
It can be set to automatically turn off the LCD display after a period of inactivity, and resume after touching.
PC software: Can realize remote control and real-time monitoring
Print Function: Can be connected to a printer to print real-time temperature and humidity or real-time operating curves
Communication Method: With 1 USB interface, 1 LAN interface
Software playback function
Historical data can be played back and converted into ACCESS or EXCEL format files. The controller can store 600 days of historical data (under 24-hour operation), which can be viewed directly in the machine. The test program is compiled through the PC special software and saved to the U disk, and then the test program is called out from the U disk and stored in the controller; the program in the controller can also be transferred to the U disk. The test program is realized through the network interface The two-way transmission between the PC and the controller directly transmits the recorded test curves and data. The test curve data recorded in the controller can be transferred to the U disk. Directly through PC special software
Connect to display and print test data and curves. Or convert the recorded data into an Access data file read by Microsoft office.
Safety protection measures:
The equipment provides the following safety protections, and sounds and lights alarms when a fault occurs:
1. Three-phase power supply lack phase protection; 2. No fuse switch protection;
3. Heater short circuit protection; 4. Blower motor overload protection;
5. High pressure protection of refrigeration system; 6. Overload protection of compressor;
7. Protector to prevent dry burning; 8. Three-color lamp operation instruction;
9. Safe and reliable grounding protection (to prevent accidental electric shock during loading process);
10. Independent over-temperature protection (when the temperature of the laboratory exceeds the set value, the heating power is cut off and an audible and visual alarm is issued).
Equipment usage conditions:
Power requirements: AC 3ψ5W 380V 50HZ;
Ambient temperature: 5~38℃, humidity: <90%R.H;